1240 Calibration resistor

Low capacitance and low inductance design, suitable for direct current and technical frequencies, high stability < ± 0.01 % over years

Product Inquiry

    Product comparison

      • Features

        • Range 10 μΩ ... 100 kΩ
        • Low capacitance and low inductance design
        • Suitable for direct current and technical frequencies
        • High stability < ± 0.01 % over years
      • Product data

        Model 1240
        Resistance value

        10 μΩ
        25 μΩ
        50 μΩ
        100 μΩ
        200 μΩ
        500 μΩ
        1 mΩ
        2 mΩ
        5 mΩ
        10 mΩ
        20 mΩ
        50 mΩ
        100 mΩ
        200 mΩ
        500 mΩ
        1 Ω
        2 Ω
        5 Ω
        10 Ω
        20 Ω
        50 Ω
        100 Ω
        200 Ω
        500 Ω
        1 kΩ
        2 kΩ
        5 kΩ
        10 kΩ
        20 kΩ
        50 kΩ
        100 kΩ

        Tolerance

        ± 0.02 %
        ± 0.03 %
        ± 0.05 %
        ± 0.1 %
        ± 1 %

        Temperature coefficient ≤ ± 10 ppm/K
        Storage stability < ± 0.01 % over years
        Power 2.5 W
        Range of operating temperature 0 ... +23 ... +40 °C
        Specific characteristics

        • Low capacitance and low inductance design
        • Suitable for direct current and technical frequencies
        • High stability < ± 0.01 % over years
      • Description

        The 0.02 class calibration resistors excel in their modern design and small mechanical dimensions. Their ruggedness also ensures a long life. Calibration resistors of the 1240 series are used wherever very constant operating standards are required. 

        Typical main areas of application therefore include:

        • When normal resistors prove to be too large scale
        • For test and calibration of resistance measurement devices
        • For tests on electrical temperature measuring equipment
        • For laboratory setup of a Wheatstone bridge
        • As shunt resistor for accurate current measurement
        • As part of standard equipment in research laboratories
        • For a large part of measurements in calibration laboratory.

        A test certificate according to ISO 9000 with detailed technical data is included in the scope of delivery of these high-quality calibration resistors.

        DKD/DAkkS Calibration Certificate

        The calibration laboratory D-K-15141-01-00 at burster praezisionsmesstechnik gmbh & co kg is supervised by DAkkS (Deutsche Akkreditierungsstelle GmbH) according to ISO 17025. 
        It can prove its status by a certificate and is authorized to issue calibration certificates with the DAkkS logo and with the DKD logo (Deutscher Kalibrierdienst). 
        These calibration certificates are internationally approved by multilateral contracts.

        Manufacturer Calibration Certificate

        Please refer to DKD/DAkkS Calibration Certificate but with reduced accuracy. The calibration resistors can also be delivered with a manufacturer calibration certificate. It confirms the traceability of the used secondary voltage and resistance standards to the national standards according to DIN ISO 9000ff and is guaranteed by our certified calibration laboratory (D-K-15141-01-00).

      • Data Sheet & Downloads

        Data sheet


      burster in Germany

      burster praezisionsmesstechnik gmbh & co kg
      +49 7224645-0
      infoburster.com
      www.burster.com

      Company-Information Company-Information

      Germany

      burster praezisionsmesstechnik gmbh & co kg

      Talstr. 1-5
      76593 Gernsbach
      Germany

      Fon: +49 7224645-0
      Fax: +49 7224645-88
      infoburster.com
      www.burster.com

      Trade fairs

      Wire
      15.04. - 19.04.24 - Düsseldorf, 40474, Germany  

      The Battery Show
      18.06. - 20.06.24 - Stuttgart, 70269, Deutschland  

      ×
      © 2024 // burster präzisionsmesstechnik gmbh & co kg
      nach oben