DIGISTANT® 4463High-precision calibration source for voltage, current, thermocouples, RTD, resistance and frequency

Smart, convenient and highly precise! The traceable solution for your calibration requirements enables fully automatic test and simulation sequences for measuring equipment and transmitters. Compared to other calibration sources, the device impresses with an outstanding error limit of just 0.002 % across the entire voltage range. Delivery includes DAkkS certificate - for maximum precision and reliability.

Add to watch list

Productinformation

DC voltage simulation
  • ±300 mV
  • ±3 V
  • ±30 V
  • ±100 V
Measurement accuracy voltage simulation
  • ±0,002 %
DC voltage measurement
  • No
DC current simulation
  • ±25 mA
  • ±50 mA
Measurement accuracy current simulation
  • ±0.005 % + 1 µA
DC current measurement
  • No
Thermocouples simulation
  • Yes
RTD simulation
  • Yes
Interfaces
  • RS232 (D-SUB9)
  • USB slave port (model B)
  • Ethernet Western socket (RJ45)
Supply voltage
  • 230 V ±10 % / 47 ... 63 Hz
  • 115 V ±10 % / 47 ... 63 Hz
Specific characteristics
  • DC voltage ±100 nV ... ±100 V (error limit from 0.002 %)
  • DC current ±100 nA ... ±50 mA (error limit ±0.005 % + 1 µA)
  • 12 thermocouple types (error limit from 0.1 K)
  • 32 autom. Ramp functions per measured variable with 100 values each
  • LabView driver for software integration

  • High-precision calibration source
  • Request for quotation Calibration services at attractive conditions

  • 4463
  • 4463 Firmware Uploader

  • 4463 Firmware Uploader
  • Instrument drivers and programming examples (X86)
  • Version information Instrument driver (X86)
  • Instrument drivers and programming examples (X64)
  • Version information Instrument driver (X64)

The DIGISTANT® model 4463 is a high-precision calibration source that impresses with its versatility and accuracy. The device is therefore supplied with a DAkkS certificate. Compared to other calibration sources, it offers a better error limit of 0.002 % over the entire voltage range.

In order to achieve a consistently high level of quality and comply with standards and directives, a wide variety of measuring devices and process instruments must be calibrated regularly. The DIGISTANT® model 4463 offers a wide range of the functions required for this. Inaccuracies caused by the connecting cables can be compensated via sense lines in 4-wire technology.

All important information on the set parameters and the accuracy achieved is clearly shown on the high-resolution colour display. The dynamic menu navigation is intuitive. Range selection is automatic or manual. Values can be entered precisely using the numeric or cursor keypad.

The device can be controlled via the Ethernet, USB or RS232 interface. LabView drivers for software integration and SCPI commands are provided free of charge. 32 ramp functions per range with 100 value/time sequences can be saved and started automatically.

Scales including ITS90 and IPTS68, constant or external cold junction type, are selectable for thermoelectric voltages. An optionally available external Pt100 reference junction with calibration data considered in the device leads to even lower uncertainties in the measuring chain.

  • Mounting set for 19"/5U control cabinet installation
  • External reference junction for thermocouple model 4485-V001

Video

  • Automatic sequence function (ramp function)
  • DC voltage up to ±100.0000 V
  • DC current up to ±50.0000 mA 

Task

For 100% testing of relays, they have to be energized gradually via a precision voltage source. Two switching moments are determined and then documented. If the measured value is within the predefined range, the specimen is found to be OK.

 Specific requirement

  • Automatic sequence function (ramp function)
  • ± switching
  • Interface for converting and comparing data

solution

The requirements of the measuring task are implemented using the DIGISTANT® 4463. The user-programmable sequence function (ramp function) enables a precise current/time simulation. Once the customer-specific maximum limit is reached, the current is reversed (±) and the ramp runs backwards to find the second switching moment. The interface makes it possible to track the values of the applied voltage and document the switching moments. The relays can be sorted based on these values.

Branchen
Electronics / Electr. Engineering

Extras

Calibrate

Customised and tailored to the diverse requirements placed on quality systems in your industry.

Learn more
Compensation

We ensure that the output signal corresponds to its specified value in the unloaded state and at nominal load.

Learn more
Implementation

We support you with on-site implementation so that you can start production quickly and reliably.

Learn more
Repair

Does your burster product need to be repaired? We'll make sure it works again quickly and easily.

Learn more